Guralp Systems Limited

Previous Next

Introduction Installing CMG-5U sensors Calibrating CMG-5U sensors Pinouts and cables Specifications Revision history

Calibrating CMG-5U sensors

The 5U accelerometer is supplied with a comprehensive calibration document, and it should not normally be necessary to calibrate it yourself. However, you may need to check that the response and output signal levels of the sensor are consistent with the values given in the calibration document.

Absolute calibration

The sensor's response (in V/m/s2) is measured at the production stage by tilting the sensor through 90 ° and measuring the acceleration due to gravity. Local g at the Güralp Systems production facility is known to an accuracy of 5 digits.

Relative calibration

In addition to the response of the sensor, several other variables are calibrated at the production stage. Using these values, you can convert directly from counts (as measured in Scream!) to acceleration values and back. You can check any of these values by performing calibration experiments.

Güralp sensors and digitizers are calibrated as follows:

In this diagram a Güralp digitizer is being used to inject a calibration signal into the sensor. This can be either a sine wave or step function, depending on your requirements. As well as going into the sensor, the calibration signal is returned to the digitizer on one of the 4 Hz auxiliary (Mux) channels. The calibration signals and sensor output all travel down the same cable from the sensor to an analogue input port on the digitizer.

The signal injected into the sensor gives rise to an equivalent acceleration (EA on the above diagram) which is added to the measured acceleration to provide the sensor output. Because the injection circuitry can be a source of noise, a Calibration enable line from the digitizer is provided which can disconnect the calibration circuit when it is not required. Depending on the factory settings, the Calibration enable line must be held either high (+5 to +10 V) or low during calibration: this is given on the sensor's calibration sheet.

The equivalent acceleration corresponding to 1 V of signal at the calibration input is measured at the factory, and can be found on the 5U calibration sheet. The calibration sheet for the digitizer documents the number of counts corresponding to 1 V of signal at each input port. The sensor transmits the signal differentially, over two separate lines, and the digitizer subtracts one from the other to improve the signal-to-noise ratio by increasing common mode rejection. As a result of this, the sensor output should be halved to give the true acceleration.

All sensors are tuned at the factory to produce 1 V of output for 1 V input on the calibration channel. For example, a sensor with an acceleration response of 0.25 V/m/s2 should produce 1 V output given a 1 V calibration signal, corresponding to 1/0.25 = 4 m/s2 = 0.408g of equivalent acceleration.

The following section explains how to calibrate 5U sensors using a DM24 series digitizer and a computer running Güralp Systems Scream! software.

Using a DM24 series digitizer for calibration

Input calibration signal strength (MB)

9856 counts

Returning signal strength (Z3)

704512 counts

Sensor acceleration response

0.254 V/m/s2

Equivalent accel. from 1V calibration

1.968 m/s2

Digitizer input port sensitivity

3.507212 µV/Bit

Mux channel sensitivity

255.9253 µV/Bit

If you prefer, you can inject your own signals into the system at any point (together with a Calibration enable signal, if required) to provide independent measurements, and to check that the voltages around the calibration loop are consistent. For reference, a DM24-series digitizer will generate a calibration signal of around 16000 counts / 4 V when set to 100% (sine-wave or step), and around 10000 counts / 2.5 V when set to 50%.

Previous Next

Introduction Installing CMG-5U sensors Calibrating CMG-5U sensors Pinouts and cables Specifications Revision history