Chapter 8. Specifications
Outputs and response | Low gain output options | 2g, 1g, 0.5g, 0.1g |
Corresponding high gain outputs | 0.2g, 0.1g, 0.05g, 0.01g | |
Dynamic range at 2 g | standard | |
Dynamic range, 0.005 – 0.05 Hz | > 140 dB | |
Dynamic range, 3 – 30 Hz | > 127 dB | |
Standard frequency band | DC – 100 Hz (–3dB point) | |
Optional low-pass corner | 50, 100 or 200 Hz | |
Linearity | 0.1 % of full scale | |
Cross-axis rejection | 0.001g / g | |
Calibration controls | Open-loop response | pin on connector |
Closed-loop response | pin on connector | |
Step function response | may be added to open- and closed-loop calibrations | |
External inputs | Sine-wave, step, or pseudo-random | |
Physical | Lowest spurious resonance | 450 Hz |
Operating temperature range | –20 to +70 °C | |
Pressure jacket material | hard anodised aluminium | |
Power / signal connector | Mil-spec connector on sensor housing (02E-14-19P) | |
Borehole diameter | 82 – 120 mm | |
Weight | 2270 g | |
Power | Current at 12 V DC | 8 mA per axis |
Specifications are subject to change without notice. Please verify requirements at time of order.