Guralp Systems Limited

Previous Next

Introduction First encounters Installing the 6TD Configuration with Scream! Calibrating the 6TD Command line interface Updating the 6TD  Connector pin-outs  Specifications Revision history

Installing the 6TD

Handling notes

Although it has a rugged design, the 6TD is still a sensitive instrument, and is easily damaged if mishandled. If you are at all unsure about the handling or installation of the device, you should contact Güralp Systems for assistance.

All parts of the 6TD are waterproof.

Connections

The instrument

The 6TD's output connectors are all located on the sensor lid. The sensor can be supplied with a number of options, so not all the connectors may be present on your instrument.

All 6TD instruments have a 19-pin mil-spec connector which carries power, data and GPS signals. The supplied breakout box (see below) provides individual connectors for these, or you can make up your own cable if you prefer.

The 6TD may also have connectors for the FireWire, Ethernet or Wi-Fi interfaces.

The connectors for the FireWire and Ethernet options are both six-pin mail-spec plugs. If the labelling has become illegible for any reason, they can be distinguished by noting that the FireWire connector lies to the East of the instrument and the Ethernet connector to the North-East, as in the picture below (which shows an early, unlabelled unit).

The breakout box

In addition to the cable to the instrument, which is moulded into the case, the breakout box provides

You may need to attach a suitable connector to the power cable provided. The 6TD draws a nominal current of 75 mA from a 12 V supply when in use; thus, using a 12 V, 25 Ah sealed heavy-duty lead-acid battery, you should expect the instrument to operate for around a week without recharging.

Installation notes

For the best possible results, a seismometer should be installed on a seismic pier in a specially-built vault, where conditions are near perfect. Here, wave-trains arriving at the instrument reflect very well the internal motion of subsurface rock formations. However, this is not always feasible. For example,

In these situations, the seismometer and its emplacement need to be considered as a mechanical system, which will have its own vibrational modes and resonances. These frequencies should be raised as high as possible so that they do not interfere with true ground motion: ideally, beyond the range of the instrument. This is done by

In temporary installations, environmental factors are also important. The sensor needs to be well protected against

This can be done by selecting a suitable site, and placing the instrument in a protective enclosure. An open-sided box of 5 cm expanded polystyrene slabs, placed over the instrument and taped down to exclude draughts, makes an excellent thermal shield.

After installation, the instrument case and mounting surface will slowly return to the local temperature, and settle in their positions. This will take around four hours from the time installation is completed.

Installing in vaults

You can install a 6TD in an existing seismic vault with the following procedure:

Installing in pits

For outdoor installations, high-quality results can be obtained by constructing a seismic pit.

Depending on the time and resources available, this type of installation can suit all kinds of deployment, from rapid temporary installations to medium-term telemetered stations.

Ideally, the sensor should rest directly on the bedrock for maximum coupling to surface movements. However, if bedrock cannot be reached, good results can be obtained by placing the sensor on a granite pier on a bed of dry sand.

Other installation methods

The recommended installation methods have been extensively tested in a wide range of situations. However, past practice in seismometer installation has varied widely.

Some installations introduce a layer of ceramic tiles between a rock or concrete plinth and the seismometer (left):

However, noise tests show that this method of installation is significantly inferior to the same concrete plinth with the tiles removed (right). Horizontal sensors show shifting due to moisture trapped between the concrete and tiling, whilst the vertical sensors show pings as the tile settles.

Other installations have been attempted with the instrument encased in plaster of Paris, or some other hard-setting compound (left):

Again, this method produces inferior bonding to the instrument, and moisture becomes trapped between the hard surfaces. We recommend the use of fine dry sand (right) contained in a box if necessary, which can also insulate the instrument against convection currents and temperature changes. Sand has the further advantage of being very easy to install, requiring no preparation.

Finally, many pit installations have a large space around the seismometer, covered with a wooden roof. Large air-filled cavities are susceptible to currents which produce lower-frequency vibrations, and sharp edges and corners can give rise to turbulence. We recommend that a wooden box is placed around the sensor to protect it from these currents. Once in the box, the emplacement may be backfilled with fresh turf to insulate it from vibrations at the surface, or simply roofed as before.

By following these guidelines, you will ensure that your seismic installation is ready to produce the highest quality data.

Rapid installation

The 6TD is specially designed for rapid installation, and may be fully installed in a few hours. This section details a method of deploying 6TD instruments with the minimum of additional equipment. This is recommended for situations where seismic instrumentation needs to be installed very quickly, e.g. to study a resumption of volcanic activity, or where difficulty of access to the site prevents you from constructing a full seismic pit. You should always construct a pit if possible (see section 3.5, page 20), since the data produced will be of significantly higher quality.

Recovery

Care should be taken when recovering the 6TD, since tapping or banging it can cause damage to the sensors inside. The following instructions assume that you have installed the instrument following the steps above.

Setting up the Ethernet interface

CMG-6TD instruments with Ethernet features installed use an embedded Lantronix X-Port module to provide the network interface. This module can be configured using a built-in Web server.

Before you can access the Web server, however, you will need to assign the device an IP address. This can be done using Lantronix' DeviceInstaller utility for Microsoft Windows, or using a DHCP server. You will need a PC with a network interface installed.

Using DeviceInstaller

Using DHCP

If you cannot install DeviceInstaller on your PC, or do not wish to, you can also get access to the 6TD using a standard DHCP server. In most cases you will need to have administrative privileges to do this.

Configuration with the Web interface

Once you have access to the X-Port's Web interface, you can configure it with its proper settings.

Setting up wireless networking

CMG-6TD instruments with wireless features installed use an embedded Lantronix Wi-Port module to provide the network interface. This module can be configured using the DeviceInstaller utility for Microsoft Windows, or using a DHCP server. You will need a PC with a wireless card installed.

You may find it easiest to gather together all the Wi-Fi hardware before taking it into the field, and configuring it from a local wireless-enabled PC.

CMG-6TD instruments with the wireless networking option also have an ETHERNET port for attaching to a wired network. You can switch between the wired and wireless interfaces using DeviceInstaller.

There are two types of wireless network topology supported by the Wi-Port.

Ad hoc networks are easy to set up, but they are only suitable with a small number of hosts. In seismic networks, infrastructure mode is normally preferred, since sensors do not need to communicate with each other.

Using DeviceInstaller in an infrastructure network

Using DHCP in infrastructure mode

If you cannot install DeviceInstaller on your PC, or do not wish to, you can also get access to the 6TD using a standard DHCP server. In most cases you will need to have administrative privileges to do this.

Using “ad hoc” mode

If you do not have a wireless router or access point, you can configure your computer to set up an ad hoc wireless network when the 6TD comes within range.

To configure Windows XP to set up an ad hoc wireless network:

Configuration with the Web interface

Once you have access to the Wi-Port's Web interface, you can configure it with its proper settings.

Installing wireless hardware

The small antenna supplied with the 6TD is adequate for initial testing or temporary installations with an access point within 50 m of the instrument.

To send data over a larger distance, or if the line of sight between the antenna and the access point is blocked, you will need to use a larger and more powerful antenna.

In infrastructure mode, you can reduce the power requirements by using a directional antenna pointed at the location of the access point. The access point does not need to be permanently present. For example, you could set up an array of 6TD instruments with antennas pointed towards a prominent natural feature with line of sight to all the instruments, and access them all from this location using a laptop PC.

Configuring the built-in digitizer

Autonomous 6TD installations will need to be configured before deployment. You can do this either

Both methods provide full access to the configuration options of the built-in digitizer.

In particular, 6TD can operate in a number of filing modes. These modes determine whether the unit stores data in its on-board Flash memory, sends it over the serial link in GCF format, or does some combination of these. See section 4.2, page 64, for more details.

Downloading data over FireWire

The easiest way to download data over Firewire is to connect a suitable disk to the FireWire port of the 6TD and power cycle the instrument.

If you have ordered a 6TD with the powered FireWire option, you can attach the disk directly to the 6TD with no additional connections. Otherwise, you will need to connect the disk to a power source through the supplied adapter.

When the sensor restarts, it will automatically the disk and flush all new data to it.

If you do not want to restart the instrument, you can also flush data to disk manually:

Reading 6TD disks

The 6TD uses a special disk format, DFD, for recording data. This format is also used by other Güralp digitizers such as the DM24.

You can read this data into a PC using Scream! or the gcfxtract utility, which are freely available from the Güralp Systems Web site.

Note: The DFD format is not the same as that used by the Güralp Systems DCM data module, which uses a FAT32-compatible journalling file system.

Güralp Systems can provide fully-tested disks with FireWire and USB connectors. Alternatively, a third-party FireWire disk may be used (although compatibility is not guaranteed).

To read a disk using gcfxtract:

gcfxtract will scan the disk and display all the streams it finds in the selection area below. For each stream, the Stream ID and the number of blocks found are shown.

This operation requires roughly 12 Mb of available memory for every Gb of space on the disk. If you have a very large disk, your computer may have to use its hard disk to make enough space. This will slow down scanning considerably.

You can check or uncheck all of the boxes using the Select all and Unselect all buttons. Clicking Invert checks all unchecked boxes, and unchecks all checked boxes.

For data streams, if there is a gap in the data, gcfxtract will start a new file anyway.

Status streams are also saved in in multiple files, but have a default length of 24 hours. To change this, alter the value under Status File Duration (hours).

You can also read disks with Scream!. This allows you to view data in the process of being transferred, but is slightly slower, because Scream! does not read data in strict order. To read a disk with Scream!:

Receiving data in Scream!

There are several ways a 6TD instrument can connect to Scream!:

To connect to a 6TD over the network:

Previous Next

Introduction First encounters Installing the 6TD Configuration with Scream! Calibrating the 6TD Command line interface Updating the 6TD  Connector pin-outs  Specifications Revision history