Guralp Systems Limited
MAN-BHO-0005 - CMG-5TB Operator's guide.

Chapter 3. First encounters

3.1 Unpacking and packing

The 5TB seismometer is delivered in a single transportation case, which is specifically designed for the 5TB. This packaging should be reused whenever you need to transport the sensor. Please note any damage to the packaging when you receive the equipment, and unpack on a clean surface. The package should contain:

3.2 Handling notes

The 5TB is a sensitive instrument, and can be damaged if mishandled. It will not stand vertically upwards without support, and should not be operated until it has been securely installed in a borehole casing. If you are at all unsure about the handling or installation of the device, you should contact Güralp Systems for assistance.

The 5TB is delivered as a single package, with its sections (including hole lock and magnetometer units, if ordered) already joined together. You should not need to disassemble it.

3.3 Control units

Once it has been installed, the 5TB is operated from the surface through various control units. All the 5TB's functions can be accessed through one or other unit. Most can be removed from the site once the instrument is ready for use.

Some of these control units are optional and may not have been supplied with your installation. Their functions can be duplicated either by applying voltages directly to control lines (see appendixes for pin-out information) or through a connected Güralp digitizer such as the DM24. The DM24 digitizer is able to pass commands to the instrument from an Enhanced Acquisition Module (EAM) or a computer running Güralp Systems' Scream! software, allowing you to access all of the instrument's functions remotely.

3.3.1 The surface control unit

This box should be installed in a surface enclosure, if you are not connecting the 5TB directly to a digitizer. It provides connections for the various elements of the sensor system.

If you are using a down-hole DM24 digitizer, it will be supplied with its own surface control unit, which differs from the one described here. See the DM24 operator's guide for more details.

The borehole control unit is supplied with a 3 metre cable as standard. This can be extended up to 100 metres without compromising signal quality. The sensor uses amplifiers with high common mode rejection to ensure the signal to noise ratio is maintained over this distance. Individually shielded twisted-pair cabling must be used for the sensor outputs, control lines and power supply. If you need to make up a suitable cable, you should confirm the cable type with Güralp Systems.

3.3.2 The hand-held control unit

This portable control unit provides easy access to the seismometer's control commands, as well as displaying the output velocity and mass position (i.e. acceleration) on an analogue meter.

This HCU is based on that used for the Güralp 40T, so some of the markings are not relevant to the 5TB.

3.3.2.1 Connections

The HCU provides

3.3.2.2 Signal meter

The upper section of the HCU contains a simple voltmeter for monitoring various signals from the instrument.

3.3.2.3 Calibration and control

You can calibrate a 5TB sensor through the HCU by connecting a signal generator across the yellow and green CALIBRATION SIGNAL inputs and setting the adjacent switch to ON. The sensor's response can now be monitored or recorded, and calibration calculations carried out.

The section of the HCU below the calibration lines controls the motors which adjust DC offsets within the instrument. You can use this section to trim DC offsets within the 5TB: see section 3.4 on page 13 for details.

3.3.2.4 Banana plugs

The remainder of the HCU provides useful connections for each of the signal lines from the instrument, for attaching to your own equipment as necessary.

3.3.3 The High-gain/low-gain break-out box

This box separates the high-gain and low-gain outputs of the instrument and routes them to two separate recorder connections, so that both outputs can be digitised at once, allowing the operator to take advantage of the full dynamic range of the instrument.

The six pins which carry the main differential outputs on the high-gain connector are coupled to the high-gain outputs from the instrument. The same six pins on the low-gain connector are coupled to the low-gain outputs from the instrument. With the exception of the power input lines, the other signals are connected as normal so that either recorder connection can be used alone. The power input lines are not connected, so as to avoid accidentally connecting the power outputs of two digitisers together, as described below.

You can connect two separate digitisers as shown:

Note: Many digitisers have power outputs so that they can power sensors directly. To avoid shorting together the power outputs of the two digitisers, with consequent risk of damage or even fire, the relevant power input pins on the break-out box recorder connectors are not wired internally. Power for the instrument must be provided through the separate power input connector on the break-out box.

If you have a six-channel digitiser with a switched/monitored power output, such as a Güralp DM24S6EAM or Güralp Affinity, you can connect both of the instrument's outputs simultaneously and use the switched outlet for the instrument, as shown:

3.4 Adjusting DC offsets

Although the 5TB masses do not require centring, you may find that a component outputs a constant DC signal. Inside the 5TB, three motors drive potentiometers which can be used to zero these offsets. The offset circuitry is designed so that each component keeps its full dynamic range whatever the position of the potentiometer.

The easiest way to zero DC offsets is through the Hand-held Control Unit:

If you prefer, you can zero the offsets manually. The X, Y and Z Motor lines from the sensor's output, together with the Motor Return line, are brought out on the HCU and RECORDER connectors of the sensor control unit. To operate a motor, apply 4 V between the X, Y or Z line and the Return line. Reverse the polarity to make the motor turn in the opposite direction. If you do not have a source of 4 V, you can use a 12 V source in series with a 47 Ω resistor.

3.5 Operating the hole lock

The hole lock, if fitted, can be extended and retracted using the hole lock control unit:

Warning: The hole lock may be using high-voltage mains (outlet) power. Observe suitable precautions.

3.5.1 Engaging the hole lock

To extend the jaw of the hole lock:

3.5.2 Disengaging the hole lock

To retract the jaw of the hole lock:

3.5.3 Manual operation

If you prefer, you can operate the hole lock by applying voltages directly: