Guralp Systems Limited
MAN-ELP-0110 - Pre-amplifier for GS13- Operator's Manual

Chapter 4. Connector pin-outs

4.1 Instrument connector

This is a standard 26-pin “mil-spec” socket, conforming to MIL-DTL-26482 (formerly MIL-C-26482). A typical part-number is 02E-16-26S although the initial “02E” varies with manufacturer.

Suitable mating connectors have part-numbers like ***-16-26P and are available from Amphenol, ITT Cannon and other manufacturers.

Pin

Function

Pin

Function

A

Coil +

P

not connected

B

Coil -

R

Calibration +

C

not connected

S

not connected

D

not connected

T

not connected

E

not connected

U

not connected

F

not connected

V

not connected

G

not connected

W

not connected

H

Calibration -

X

not connected

J

not connected

Y

Calibration return

K

not connected

Z

not connected

L

not connected

a

Calibration shield and casing

M

not connected

b

not connected

N

Coil shield

c

not connected

Wiring details for the compatible plug, ***-16-26P, as seen from the cable end (i.e. when assembling).

4.2 Digitiser connector

This is a standard 26-pin “mil-spec” plug, conforming to MIL-DTL-26482 (formerly MIL-C-26482). A typical part-number is 02E-16-26P although the initial “02E” varies with manufacturer.

Suitable mating connectors have part-numbers like ***-16-26S and are available from Amphenol, ITT Cannon and other manufacturers.

Pin

Function

Pin

Function

A

Signal +

P

Calibration signal

B

Signal -

R

Calibration enable

C

not connected

S

not connected

D

not connected

T

not connected

E

not connected

U

not connected

F

not connected

V

Active-high select *

G

not connected

W

not connected

H

not connected

X

not connected

J

not connected

Y

Logic ground

K

not connected

Z

not connected

L

not connected

a

not connected

M

not connected

b

Power supply – zero volts

N

Signal ground

c

Power supply - positive

Wiring details for the compatible socket, ***-16-26S, as seen from the cable end (i.e. when assembling).

* DM24 digitisers normally provide active-low control outputs but the “calibration enable” input of the pre-amplifier is active-high. CTBTO models can be switched to active-high mode by connecting pin V to logic ground (pin Y). The Güralp ELP-0110 provides this connection internally.