Chapter 13. Appendix C – Deployment check-list
This deployment checklist is provided for information only. Refer to Section 9 for full details of the deployment procedure.
Aquarius S/N: _______________
Acoustic modem address: _________________
Acoustic modem UID:________________
Deployment location: Latitude _____________________, Longitude:___________________
Deployment date:_____/_______/____________
Operation | Check |
Fully charge the batteries inside the Aquarius (see Section 5.4). | |
Screw the charger pressure cap back in. | |
Attach the ballast to the Burn-Wire system (see Section 8.2 page 65). | |
Connect the Seacon Hummer series connector on the Aquarius and the Ethernet and console connectors to the Deck Unit. | |
Remove the magnetic off plug and fit the vent cap with the O-ring and the sacrificial anode / If the cap with the O-ring is already in place and the system is already on, switch PoE on, reset it using the System Reset button in the OBS On Deck tab of OBS Command & Control widget. | |
In the Deck Unit, switch the PoE on. | |
Check PTP lock either from the LPC web page (Status tab, PTP status section, PTP stability cell) or from the OBS Command & Control widget /Pre-Deploy tab / Timing Status. If No Master is shown in the above fields, check PTP network configuration in the LPC (see Section 7.2.2) or GPS fix in the PTP server. | |
Check whether recording of non FR channels for Sensor 0 and Sensor 1 are disabled in the LPC web page / Data Record tab (see Section 7.2.4). | |
Select all the FR channels on Sensor1 to be recorded from the LPC web page / Data Record tab. | |
If changes were made in the LPC web page / Data Record tab, reboot the LPC. | |
In the LPC web page / Trigger tab, set as trigger source S1RatioZFR and choose appropriate values for STA/LTA parameters. Set the Whalesong Sender as trigger destination. Set the threshold to 0 (see Section 7.2). This will disable triggering until the OBS is on the seabed. At that point this parameter can be modified via acoustics, if a buoy unit is present. |
|
Download the Dataless file from the Storage tab of the LPC web page. | |
Set the bottom acoustic modem address in the Network tab of the LPC web page (Local modem address) and take note of it. In case of an installation with a buoy unit on the surface, set also the surface acoustic modem address (Remote modem address in the LPC web page, see Section 7.2.6). | |
In the LPC web page, Network tab, set temporarily the acoustic modem Power level & gain to Suitable in air (see Section 7.2.6) | |
Test the Burn-Wires via acoustic command from the Recovery tab in the OBS Command & Control widget (see Section 7.1.3.5). | |
In the LPC web page, Network tab, set the acoustic modem Power level & gain to a depth suitable for the specific installation (see Section 7.2.6). | |
Quick format the SD cards either from the LPC web page (Storage tab) or from the Pre-Deploy tab of the Discovery OBS Command & Control widget (Format button). | |
Check live data from sensors in the Discovery Live View (see Section 7.1.2). | |
Test Burn-Wires from ULPD using the Discovery OBS Command & Control widget, OBS On Deck tab, Test Burn Wire button (see Section 7.1.3.2). | |
Test the satellite tracker and LED strobe activating them from the Discovery OBS Command & Control widget, OBS On Deck tab, Test Satellite button. | |
Set the time for the backup Burn-Wires that would be activated by the ULPD timer. Use the Discovery OBS Command & Control widget, OBS On Deck tab for this task (see Section 7.1.3.2). | |
If the Offset in OBS Command & Control / OBS On Deck tab / Timing Status is not 0ms force a time synchronization using the button Force Sync. | |
Wait until the message Good to go is displayed in the OBS On Deck tab in OBS Command & Control (see Section 7.1.3.2). | |
DEPLOY the Aquarius either from the LPC web page (see Section 7.2.7) or from the Discovery OBS Command & Control widget, OBS On Deck tab, Deploy button. | |
Switch off the PoE in the Deck Unit and unplug the Seacon Hummer series connector. | |
Plug the Seacon Hummer series dummy cap. | |
Fit the central block of buoyancy and mount the lifting frame. | |
Deploy the Aquarius in the water. | |
Range the Aquarius while is descending towards the seabed. For this task, you could either use the Modem Ping button in the OBS Command & Control widget / Post-Deploy tab or the OBS location widget (see Section 9.2.2.1). | |
Click on the Status button in the OBS Command & Control widget / Post-Deploy tab to show the Aquarius depth (inferred from the Keller pressure sensor data) and other SoH info. | |
When the Aquarius will have reached the seabed, wake up the LPC asking for the Aquarius ID or for the SEED ID using the OBS Command & Control widget, Post-Deploy tab. | |
Force a centring via acoustics using the button Perform Recentre in the Post-Deploy tab of the OBS Command & Control widget (see Section 43). | |
Check re-centring status using the button Get Recentring Status in the Post-Deploy tab of the OBS Command & Control widget. | |
When centring is completed, flush data from the ULPD to the LPC microSD cards using the Flush Data button in Post-Deploy tab of the OBS Command & Control widget. | |
Retrieve data via acoustic comms to check sensor performance (see Section 7.1.3.4). |